Optical coherence tomography images tissue in depth by low-coherence interferometry: interference occurs only when path lengths match within the coherence length, giving micron axial resolution set by the source bandwidth.
Explanation
A Michelson interferometer with a broadband source: light backscattered from different depths interferes with the reference arm only when the path difference is within the coherence length, so scanning the reference (time domain) or measuring the spectrum of the combined light (Fourier domain) yields a depth profile (A-scan). Lateral scanning builds cross-sections.
Axial resolution depends only on bandwidth, lateral resolution on the focusing optics, which decouples them. OCT is standard in ophthalmology (retinal layers) and used in cardiology and dermatology; near-infrared light gives millimetre penetration in scattering tissue.
Example 1
An OCT source has centre wavelength 840 nm and Gaussian bandwidth 50 nm. Find the axial resolution in air and in tissue ($n=1.38$).
Solution
$\Delta z=0.441\times\frac{(840\ \mathrm{nm})^2}{50\ \mathrm{nm}}\approx6.2\ \mu$m in air, $\approx4.5\ \mu$m in tissue — enough to see individual retinal layers.
Example 2
Why would a narrow-band laser be useless as an OCT source in the time-domain scheme?
Solution
Its coherence length (metres) exceeds the sample depth, so reflections from all depths interfere with the reference at every arm position; no depth gating occurs. Short coherence length is what localises the reflection.