A microscope cannot resolve details finer than about λ/2NA, because the objective must collect at least the first diffraction orders; immersion oil raises the numerical aperture NA = n sin θ.
Explanation
A grating of period $d$ diffracts its first orders at $\sin\theta=\lambda/d$; the image shows the grating only if these orders enter the objective (with oblique illumination the limit halves to $\lambda/2{\rm NA}$). For point sources Rayleigh's criterion gives $0.61\lambda/{\rm NA}$. Oil ($n\approx1.52$) raises NA up to about 1.4.
Magnification beyond $\sim500$–$1000\times{\rm NA}$ is empty: it enlarges blur, not detail. Fluorescence super-resolution methods (STED, localisation microscopy) circumvent the limit by exploiting nonlinear or switchable emitters, not by beating diffraction itself.
Example 1
An oil-immersion objective has NA = 1.40. Find the Abbe and Rayleigh limits at 500 nm.
Solution
$d_{\rm Abbe}=\frac{500}{2.8}\approx180$ nm; $d_{\rm Rayleigh}=0.61\times\frac{500}{1.40}\approx220$ nm.
Example 2
The eye resolves about 0.1 mm at 25 cm. What is the largest useful magnification for a microscope resolving 0.2 μm?
Solution
$M\approx\frac{0.1\ \mathrm{mm}}{0.2\ \mu\mathrm m}=500$, consistent with the rule $500$–$1000\times{\rm NA}$.