Confocal microscopy and optical sectioning

Database › Physics › Optics › Optical design

OPT-0069MethodUndergraduate

A confocal microscope illuminates a diffraction-limited spot and detects through a conjugate pinhole, rejecting out-of-focus light; this gives optical sectioning and a modest gain in lateral resolution.

$$\Delta x_{\rm conf}\approx0.37\frac{\lambda}{\rm NA},\qquad\Delta z_{\rm conf}\approx\frac{0.88\lambda}{n-\sqrt{n^2-{\rm NA}^2}}\ \ ({\rm FWHM})$$

Explanation

The overall point-spread function is the product of the illumination and detection PSFs, which is narrower than either and, crucially, falls off with defocus: light from planes above and below the focus is spread over the pinhole plane and mostly blocked. Scanning the spot builds images slice by slice, so thick fluorescent samples can be reconstructed in 3-D.

Two-photon excitation achieves sectioning without a pinhole because absorption $\propto I^2$ is confined to the focus, and its near-infrared light penetrates deeper into tissue.

Example 1

For an oil objective with NA = 1.40 ($n=1.518$) at 500 nm, estimate the confocal lateral and axial resolution, and compare the lateral value with the widefield $0.51\lambda/{\rm NA}$.

Solution

Lateral $0.37\times500/1.40\approx130$ nm versus widefield $\approx180$ nm. Axial $\frac{0.88\times500}{1.518-\sqrt{2.304-1.96}}\approx\frac{440}{0.93}\approx470$ nm: the focal volume is elongated along $z$.

Example 2

Why does two-photon excitation not need a detection pinhole?

Solution

Excitation rate $\propto I^2$, and the integral of $I^2$ over a plane falls as $1/z^2$ away from focus (the beam area grows as $z^2$). Fluorescence is therefore generated almost only in the focal volume, so all emitted light can be collected.